Deep Brain Stimulation Electrode Deviations are Associated With Brain Stiffness Interfaces Measured by Magnetic Resonance Elastography.
Аннотация
BACKGROUND AND
OBJECTIVES
The efficacy of deep brain stimulation (DBS) relies on accurate electrode placement. Unfortunately, electrode deviation poses a persistent problem, with most electrodes demonstrating some degree of bending. Although such bending does not always result in target deviation, an estimated 3% to 8% of patients still require revision surgery to address suboptimal electrode placement. DBS electrode deviation may occur at mechanical tissue interfaces, with denser internal capsule (IC) fibers being the most likely factor. Based on basic principles of physics, we hypothesized that the angle of a planned trajectory relative to tissue interfaces created by the IC induces deviation.
METHODS
Ten patients with Parkinson disease scheduled for DBS surgery underwent preoperative 3T magnetic resonance elastography (MRE) using synchronized external vibrations to measure brain tissue stiffness. The IC stiffness interface (ICSI) was defined as the transition between the corona radiata and IC on MRE. The rate of transition was calculated as the change in stiffness across the ICSI. Postoperative computed tomography was used to measure target deviation . The angle of approach was calculated as the angle between the planned trajectory and the normal vector to the ICSI. Pearson correlations and t -tests were performed to evaluate associations between the angle of approach and target deviation.
RESULTS
Twenty-one electrode trajectories were analyzed. The mean electrode deviation was 1.27 ± 0.63 mm. A significant correlation (r = 0.57, 95% CI [0.18, 0.80], P = .007) was found between angle of approach and target deviation, with larger angles associated with greater deviations. The rate of transition did not correlate with deviation ( P = .874).
CONCLUSION
MRE effectively quantifies in vivo brain tissue stiffness in Parkinson disease. The angle between the planned trajectory and the ICSI correlates with target deviation, supporting the hypothesis that tissue mechanics influence electrode bending. MRE has potential to quantify the likelihood of DBS electrode deviation, which could reduce revision surgeries and enhance clinical outcomes.