DOI 10.1177/11206721231202539

Медицинская библиотека · European journal of ophthalmology · 2023

Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy in localizing horizontal rectus muscle insertions.

Duan Rui, Yang Jiwen

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the viability and precision of measuring the distance from the limbus to extraocular muscle insertion using anterior segment optical coherence tomography (AS-OCT) and panoramic ultrasound biomicroscopy (UBM) before and after strabismus surgery.

METHODS

We recruited primary strabismus patients and measured the limbus-insertion distance by AS-OCT and UBM preoperatively, 2 weeks, and 1, 3, and 6 months postoperatively. Values were also measured using callipers intraoperatively before and after the planned procedures. Preoperative AS-OCT and UBM values were compared to intraoperative calliper measurements as the gold standard. Postoperative AS-OCT and UBM values were compared to the new postoperative limbus-insertion distance. The limit of agreement deemed clinically acceptable was defined as 1 mm.

RESULTS

A total of 85 horizontal muscles of 40 patients, including 48 lateral rectus muscles and 37 medial rectus muscles, were analysed. Primary muscles could be successfully detected by AS-OCT (95%) and UBM (100%). At 2 weeks and 1, 3, and 6 months postoperatively, the new rectus muscle attachment site detection rate by AS-OCT was 6%, 32%, 80%, and 89%, respectively, and that by UBM was 24%, 60%, 85%, and 93%, respectively. The Bland-Altman plots revealed better consistency in pairs of AS-OCT, UBM, and calliper measurements of primary muscles than postoperative muscles. For primary muscles, 89% of AS-OCT measurements fell within the permissible range of surgical measurements (1 mm), but this dropped to 67% at 6 months postoperatively (P < 0.001). The accuracy of UBM measurements of primary muscles was 81%, and this decreased to 59% at 6 months postoperatively (P = 0.001).

CONCLUSIONS

AS-OCT and UBM performed well in terms of imaging primary horizontal rectus muscles, but showed decreased accuracy and reproducibility in postoperative muscle measures.

Для цитирования: Duan Rui, Yang Jiwen. Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy in localizing horizontal rectus muscle insertions.. European journal of ophthalmology. 2023. DOI: 10.1177/11206721231202539.