DOI 10.1016/j.healun.2017.02.003
Berlin Heart EXCOR use in patients with congenital heart disease.
Аннотация
BACKGROUND
Management of mechanical circulatory support in children with congenital heart disease (CHD) is challenging due to physiologic variations and anatomic limitations to device placement. In this study we examine the use of Berlin Heart EXCOR in CHD patients.
METHODS
CHD patients were identified from the EXCOR Pediatric Study data set (2007 to 2010). Mortality and serious adverse events were compared between CHD and non-CHD cohorts, and predictors of poor outcomes in the CHD cohort were identified.
RESULTS
CHD was present in 29% (n = 59, 18 with 1-ventricle physiology) of all EXCOR patients (N = 204). Successful bridge (transplant or wean) was less likely in CHD patients compared with non-CHD patients (48% vs 80%; p < 0.01). Among CHD patients, no neonates, 25% of infants (30 days to 1 year) and 65% of children (>1 year) were successfully bridged. Pre-implant congenital heart surgery (CHS) and extracorporeal membrane oxygenation (ECMO) on the same admission occurred in 60% of children ≤1 year of age (83% of neonates, 50% of infants), with 8% survival. Regardless of age, patients who did not have CHS and ECMO had 61% survival. Smaller pump, pre-implant bilirubin >1.2 mg/dl and renal dysfunction were independently associated with mortality.
CONCLUSIONS
End-organ function at implant reliably predicts adverse outcomes and should be considered when making implant decisions. EXCOR use in neonates and infants with CHD should be approached cautiously. If patients have undergone pre-implant CHS and ECMO, EXCOR support may not provide any survival benefit. EXCOR support in non-infants with CHD is challenging but can be consistently successful with appropriate patient selection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.