DOI 10.1016/j.avsg.2015.12.026
Preliminary Single-Center Experience with the Bolton Treovance Endograft in the Treatment of Abdominal Aortic Aneurysms.
Аннотация
BACKGROUND
Our aim was to present our preliminary experience with the recently introduced Treovance aortic stent-graft device (Bolton Medical, Barcelona, Spain) in the treatment of abdominal aortic aneurysm (AAA).
METHODS
Eight patients underwent treatment of an infrarenal AAA (mean maximum diameter, 56.4 ± 6.8 mm) with the Treovance device. Iliac tortuosity was considered mild, moderate, or severe when ≥1 angulation of 45-90°, 1 angulation ≥90°, or ≥2 angulations ≥90°, respectively, were present.
RESULTS
Mild angulation of the infrarenal neck (10-45°) was present in 7 patients, whereas the remaining patient had severe infrarenal neck angulation (65°). Three patients had severe iliac tortuosity. Primary technical success was achieved in all but 1 patient in whom a type Ia endoleak was identified on completion angiogram. The endoleak was successfully treated with a proximal aortic cuff. A femoral access complication occurred in 1 patient. Mean follow-up was 6.8 months (range, 1-12). No device-related serious adverse events or rupture occurred during the given follow-up period. The only type II endoleak identified resolved spontaneously within 12 months.
CONCLUSIONS
The Treovance abdominal stent-graft system seems to guarantee an accurate, safe, and effective deployment in AAA even through angulated and tortuous iliac vessels. Although our preliminary results are promising, follow-up data are needed to establish the durability of this new-generation endovascular device in standard or challenging anatomies.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.