Hemodynamic studies on brain CT perfusion imaging with varied injection rates.

Hu Chun-Hong, Wu Qing-De, Hu Xiao-Yun, Fang Xiang-Ming, Zhang Tong-Hua, Ding Yi

Клиническое исследованиеСравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study aimed to determine a contrast medium injection rate that ensures both accuracy for data and safety for operation by comparing hemodynamic parameters of brain CT perfusion imaging with varied injection rates.

METHODS

Twenty-four healthy volunteers were divided into three groups based on contrast medium injection rates (4.5, 6.0, and 7.5 ml/s). For all subjects, CT perfusion scanning was started at 4 s after antecubital venous bolus of contrast media injection. A perfusion-analyzing software package was used to produce a time-density curve in the anterior cerebral artery and the superior sagittal sinus and calculate the regional cerebral blood flow (rCBF) in the gray matter and the white matter. The hemodynamic indices were compared among the three groups, and statistical analysis was carried out using the F test.

RESULTS

The time for the arterial rise to reach the peak value for the 7.5-ml/s group was only 0.2 s ahead of the initiation time for the rise of the superior sagittal sinus. The differences of rCBF in the gray matter and the white matter among the three groups were statistically significant. rCBF in the gray matter and the white matter for the 7.5-ml/s group was 52.8 ml x min(-1) 100 g(-1) . (+/-3.1) and 21.9 ml x min(-1) . 100 g(-1) (+/-2.4), respectively.

CONCLUSIONS

The use of the 7.5-ml/s injection rate can meet the prerequisite of the maximum slope model, and the resulting rCBF can be very close to that measured by positron emission tomography. Therefore, 7.5 ml/s was an ideal contrast medium injection rate.

Для цитирования: Hu Chun-Hong, Wu Qing-De, Hu Xiao-Yun, Fang Xiang-Ming, Zhang Tong-Hua, Ding Yi. Hemodynamic studies on brain CT perfusion imaging with varied injection rates.. Clinical imaging. 2007. DOI: 10.1016/j.clinimag.2007.01.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.