MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.70087

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2025

Amusement Park Rides and Cardiac Devices: Heart Dropper or Device Stopper?

Kassab Mia, Johnson Alexander, Kallas Dania, Franciosi Sonia, Bone Jeffrey, Vijayashankar Sakethram Saravu, Sanatani Shubhayan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac implantable electronic devices (CIEDs) are essential for managing cardiac conditions, but may malfunction due to magnetic fields > 10,000 mG. Roller coasters using linear induction motors (LIMs) generate magnetic fields, yet their potential for electromagnetic interference (EMI) with CIEDs is unclear. This study assesses magnetic field exposure on amusement park rides and examines healthcare provider recommendations.

METHODS

Magnetic field strength was measured using gaussmeters placed at shoulder and abdomen levels, representing pediatric CIED sites. Rides at an amusement park were tested at least four times, recording median and maximum magnetic field strengths per second throughout the ride. Magnetic field strengths were compared between rides with health advisory messages (HAMs) and without (NHAMs). A survey was distributed to the Pediatric and Congenital Electrophysiology Society (PACES) and the Canadian Council of Cardiovascular Nurses to assess healthcare provider recommendations.

RESULTS

A total of 15 rides were sampled: 11 with HAMs and 4 with NHAM. The mean magnetic field strength was higher for HAM rides (2.9 mG) than NHAM rides (1.6 mG; p = 0.05). Maximum field strength was also greater in HAM rides (46.4 vs. 6.5 mG; p < 0.001), and in rides using LIMs (n = 2) compared to those using other mechanisms (211.7 vs. 7.8 mG; p < 0.001). Only 18.1% (n = 13) of healthcare providers relied on published resources for amusement park ride recommendations, while 58.3% (n = 42) advised patients to consider HAMs.

CONCLUSION

Magnetic field strengths on all rides were clinically insignificant, posing minimal EMI risk for CIED patients. Further validation and standardized guidelines are needed to inform healthcare recommendations for patients with CIEDs.

Для цитирования: Kassab Mia, Johnson Alexander, Kallas Dania, Franciosi Sonia, Bone Jeffrey, Vijayashankar Sakethram Saravu, Sanatani Shubhayan. Amusement Park Rides and Cardiac Devices: Heart Dropper or Device Stopper?. Journal of cardiovascular electrophysiology. 2025. DOI: 10.1111/jce.70087.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.