DOI 10.1136/bcr-2020-235312

Медицинская библиотека · BMJ case reports · 2020

Distortion of dual energy X-ray images by faecal masses in a child with type 1 diabetes.

Shah Nikhil, Ekbote Veena, Khadilkar Vaman, Khadilkar Anuradha

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Dual energy X-ray absorptiometry (DXA) scanning is the most common investigating modality used to assess bone mineral density (BMD). Conditions causing tissue calcification and artefacts such as metallic objects may mislead the results of the DXA scan. We present here a case of a child with diabetes where the DXA images were distorted by faecal lumps, leading to falsely elevated BMD and an error in interpretation of the DXA scans. Our case suggests that DXA software may not, at times, differentiate between bone and other high-attenuating material within the regions of interest. Thus, DXA images should also be visually examined and verified with the numeric data before report preparation in these patients.

Для цитирования: Shah Nikhil, Ekbote Veena, Khadilkar Vaman, Khadilkar Anuradha. Distortion of dual energy X-ray images by faecal masses in a child with type 1 diabetes.. BMJ case reports. 2020. DOI: 10.1136/bcr-2020-235312.