MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1526602815617477

Медицинская библиотека · Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2015

Type IIIb Endoleak With the Endurant Stent-Graft.

McWilliams Richard G, Vallabhaneni S Rao, Naik Jagjeeth, Torella Francesco, Fisher Robert K

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To report 2 cases of type IIIb endoleak with the Endurant stent-graft and postulate the cause for the events.

CASE REPORT: A type IIIb endoleak was diagnosed at open conversion for a ruptured aneurysm 4 years after implantation of an Endurant stent-graft. In the other case, the endoleak was diagnosed at angiography 4 years after the Endurant stent-graft was implanted; the stent-graft was relined. In both cases the fabric hole was in the body of the stent-graft at the level of the top of the contralateral limb.

CONCLUSION

The cause of the type IIIb endoleaks in these cases was fabric erosion likely due to interaction between the bare metal at the top of the contralateral limb and the fabric of the stent-graft body.

Для цитирования: McWilliams Richard G, Vallabhaneni S Rao, Naik Jagjeeth, Torella Francesco, Fisher Robert K. Type IIIb Endoleak With the Endurant Stent-Graft.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2015. DOI: 10.1177/1526602815617477.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.