Early experience with the new endovascular aneurysm sealing system Nellix: First clinical results after 50 implantations.
Аннотация
OBJECTIVE
In the present study, 50 EVAS procedures were evaluated in regard to primary (survival and technical success) and secondary (device-related complications) events of interest.
METHODS
The single center study was conducted from July 2013 to August 2014 with prospective collection of the clinical data. The clinical results were controlled by CT angiography and contrast-enhanced ultrasound.
RESULTS
The technical success was 98% and the 30-day mortality 4%. One (2%) patient died from multisystem organ failure and another patient from an intracranial bleeding, respectively. One patient (2%) suffered from a device-related aneurysm rupture. During early follow-up, one (2%) patient developed an endoleak type II, while three (6%) patients suffered from a partial endograft limb thrombosis. Overall, a secondary intervention was necessary in six (12%) patients.
CONCLUSIONS
With the Nellix EVAS system, a high primary technical success of 98% was achieved; one (2%) patient developed an endoleak type II which did not require secondary intervention. Those promising results are contrasted by a substantial rate of endograft limb thromboses (8%) and one (2%) intraoperative aneurysm rupture. Further studies are needed to assess the durability of the Nellix stentgraft and the occurrence of device-related complications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.