MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00540-023-03207-2

Медицинская библиотека · Journal of anesthesia · 2023

Efficacy of McGRATH®MAC videolaryngoscope blade 1 for tracheal intubation in small children: a randomized controlled clinical study.

Masui Katsuhide, Asai Takashi, Saito Tomoyuki, Okuda Yasuhisa

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Videolaryngoscopes may not be as effective in small children as they are in older children and in adults. The size 1 blade is commercially available for the McGRATH®MAC videolaryngoscope (Covidien, Medtronic, Tokyo, Japan), but its efficacy in comparison with a Macintosh laryngoscope blade 1 is not known.

AIM: The main aim of this study was to assess the efficacy of McGrath®MAC blade 1 in comparison with a conventional Macintosh laryngoscope blade 1, in children aged less than 24 months.

METHODS

Thirty-eight children aged less than 24 months were randomly allocated to one of two groups, and tracheal intubation was attempted using either a direct laryngoscope with a Macintosh blade 1 or a videolaryngoscope with a McGRATH®MAC blade 1. In another 12 children aged 2-4 years, the same comparisons were made with blade 2. The primary outcome measure was time to tracheal intubation using a size 1 blade.

RESULTS

Tracheal intubation took significantly longer with a McGRATH®MAC blade 1 (median (interquartile range): 38.0 (31.8-43.5) s) than with the Macintosh blade 1(27.4 (25.9-29.2) s) (p < 0.0001; median difference (95% CI for the median difference): 10.6 (6.4-14.0) s), mainly due to difficulty in advancing a tube into the trachea. No significant difference was observed for the size 2.

CONCLUSIONS

In small children without predicted difficult airways, time to intubate the trachea was significantly longer for a McGRATH®MAC blade 1 than a Macintosh blade 1.

CLINICAL TRIAL REGISTRATION: jRCT1032220366.

Для цитирования: Masui Katsuhide, Asai Takashi, Saito Tomoyuki, Okuda Yasuhisa. Efficacy of McGRATH®MAC videolaryngoscope blade 1 for tracheal intubation in small children: a randomized controlled clinical study.. Journal of anesthesia. 2023. DOI: 10.1007/s00540-023-03207-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.