Tibial Plateau Leveling Osteotomy Plate Contouring and Proximal Load Screw Angulation Affect Osteotomy Compression.

Mathis Karl R, Roe Simon C, Johnson Kenneth A

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the effect of contouring a tibial plateau leveling osteotomy (TPLO) plate, the associated angulation of the dynamic compression plate (DCP) hole relative to the long axis of the tibia, and angulation of the screw relative to the DCP hole on the osteotomy compression generated by load screws in a TPLO model.

STUDY

DESIGN

In vitro biomechanical study.

STUDY POPULATION: Polyoxymethylene (POM) rod and synthetic cortical bone substitute model (n = 9).

METHODS

The distal portion of a Slocum TPLO plate was attached to a horizontally positioned POM rod that was connected to a load cell. A segment of synthetic cortical bone substitute was attached to the end mount of the testing frame and adjusted to conform to the angle of the proximal portion of the TPLO plate. A 3.5 mm cortical bone screw was inserted in the proximal DCP hole and tightened to 1.5 Nm. The peak longitudinal load (N) was recorded. Screw insertion and data collection were repeated for proximal plate angles of 0-40° at 5° increments.

RESULTS

A significant increase in the compression generated was observed as the plate angle was increased from 0° to 10°. The compression ceased to significantly increase until the plate was bent more than 20°, after which a significant decrease in compression was noted. A marked reduction in the compression generated occurred at plate angles greater than 30°.

CONCLUSION

Angulation of the DCP hole and screw insertion angle can have deleterious effects on the magnitude of osteotomy compression.

Для цитирования: Mathis Karl R, Roe Simon C, Johnson Kenneth A. Tibial Plateau Leveling Osteotomy Plate Contouring and Proximal Load Screw Angulation Affect Osteotomy Compression.. Veterinary surgery : VS. 2015. DOI: 10.1111/vsu.12414.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.