MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.transci.2006.10.004

Медицинская библиотека · Transfusion and apheresis science : official journal of the World Apheresis Association : official journal of the European Society for Haemapheresis · 2007

Optimization of therapeutic procedure during LDL-apheresis - verification of the computerized model in clinical practice.

Masin Vladimir, Blaha Milan, Stransky Pravoslav, Blaha Vladimir, Cermanova Melanie, Maly Radovan, Zajic Jiri

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

UNLABELLED: LDL-apheresis is a very effective method in the treatment of resistant hypercholesterolemia when other therapies fail. To maximize the efficacy of the use of LDL-absorbers, we created a computerized model. The aim of this study is to verify it in clinical use.

PATIENTS AND

METHODS

A therapeutic technique of immunoadsorption was used, applying a pair of Lipopak columns (Pocard, Russia). Plasma was separated by a continuous-flow plasma separator (Cobe Spectra, USA); adsorption was controlled by adsorption-desorption equipment Adasorb (Medicap, Germany). 494 LDL-apheresis procedures had been performed in nine patients with primary hypercholesterolemia in the earlier, initial study; 47 other procedures (202 therapeutic cycles) were used in this verification study. The program for procedure planning uses Microsoft Excel for Windows. Complex metabolism of the LDL-cholesterol was neglected (owing to the short-time period of the procedure) and the procedure calculated as continuous filtration. The input enterer into the program includes basic patient data (mass, height, sex and initial plasma LDL level in mmol/L).

RESULTS

The results show a very good match between calculated levels and the real laboratory results in most procedures, but in some procedures we observed minor differences (0.05 mmol/L), which was caused by procedure adjustments due to technical reasons. However, some methodological and medical details must be carefully observed (initial cholesterol level, correct calculation of plasma volume, and the precise capacity of adsorbers that must not be overshot); as they influence the correct match between calculated and real results significantly.

CONCLUSIONS

Although our software uses a fairly simplified model of the LDL-cholesterol kinetics during the LDL-apheresis, it is providing a great aid in the procedure planning. It is also suitable for practical use because it only requires a few commonly used and readily available input values.

Для цитирования: Masin Vladimir, Blaha Milan, Stransky Pravoslav, Blaha Vladimir, Cermanova Melanie, Maly Radovan, Zajic Jiri. Optimization of therapeutic procedure during LDL-apheresis - verification of the computerized model in clinical practice.. Transfusion and apheresis science : official journal of the World Apheresis Association : official journal of the European Society for Haemapheresis. 2007. DOI: 10.1016/j.transci.2006.10.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.