Development of a transcranial direct current stimulation device based on current limiter for simultaneous measurement of electroencephalography: A feasibility study.

Jung Young-Jin, Lee Yun-Sung, Choi Ga-Young, Hwang Han-Jeong

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Electroencephalography (EEG) measured during transcranial direct current stimulation (tDCS) can help understand the accurate impact of tDCS on the brain, but this has been hindered due to significant inflow of tDCS-induced electrical artifacts.

OBJECTIVE

In this study, we introduce a novel tDCS device developed based on current limiter, which can prevent the generation of significant electrical artifacts.

METHODS

To verify the feasibility of our developed tDCS device, we performed simultaneous measurement of EEG during tDCS application with five different current intensities (0, 500, 1,000, 1,500, and 2,000 μA). Changes in EEG power spectral density (PSD) and correlation between the PSD of non-stimulation and tDCS condition were investigated to see whether our tDCS device can be used for simultaneous EEG recording without significant inflow of tDCS-induced electrical artifacts.

RESUTLS: The mean EEG-PSD differences between non-stimulation and tDCS condition were not significant for all stimulation current intensities. Furthermore, EEG-PSDs estimated during non-stimulation and tDCS application showed statistically high correlation for all comparison cases.

CONCLUSION

Based on the results, we could demonstrate the feasibility of our tDCS device based on current limiter for simultaneous EEG measurement, which could potentially provide a way to investigate the impact of tDCS on the brain more accurately.

Для цитирования: Jung Young-Jin, Lee Yun-Sung, Choi Ga-Young, Hwang Han-Jeong. Development of a transcranial direct current stimulation device based on current limiter for simultaneous measurement of electroencephalography: A feasibility study.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2020. DOI: 10.3233/thc-209013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.