Increased Load to Failure in Biceps Tenodesis With All-Suture Suture Anchor Compared With Interference Screw: A Cadaveric Biomechanical Study.

Smuin Dallas M, Vannatta Emily, Ammerman Brittany, Stauch Christopher M, Lewis Gregory S, Dhawan Aman

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the biomechanical characteristics of a single radially expanding all-suture anchor with an interference screw for open subpectoral long head of biceps tendon (LHBT) tenodesis.

METHODS

Eighteen fresh-frozen matched-pair human cadaveric humeri were used for this biomechanical study. The matched pair humeri were randomly assigned into 2 experimental biceps tenodesis groups: conventional interference screw (CIS) or all-suture suture anchor (ASSA). Open subpectoral LHBT tenodesis was then performed and biomechanical testing was performed using a servohydraulic test frame. A preload of 5 N was applied for 2 minutes before cyclic loading. Displacement was recorded at cycle 300 (D300) and cycle 500 (D500) and at ultimate failure. Data recorded included displacement, load to failure, displacement at failure. Paired t test was used for analysis.

RESULTS

Decreased displacement was observed for the CIS group at D300 (1.67 ± 0.57 mm vs 3.35 ± 2.24 mm; P = .04), D500 (2.00 ± 0.76 mm vs 3.87 ± 2.20 mm; P = .04), and at failure (5.17 ± 3.05 mm vs 10.76 ± 2.66 mm; P < .001). Load to failure was lower in CIS specimens (170 ± 24.5 N vs 217.8 ± 51.54 N; P = .02). Failure in each case was tendon pullout for all CIS specimens; in ASSA 6 specimens failed as the suture pulled through the tendon, 2 specimens failed by suture breakage. No difference in stiffness was observed between the 2 groups (CIS = 98.33 ± 22.98 N/m vs ASSA = 75.94 ± 44.83 N/m; P = .20).

CONCLUSIONS

Our study found that open subpectoral biceps tenodesis performed with an ASSA construct results in increased load to failure as compared with CIS. However, the CIS did demonstrate decreased displacement as compared to ASSA in this cadaveric biomechanical study.

CLINICAL RELEVANCE: ASSA and CIS at time zero provide fixation as indicated by the provider intraoperatively for LHBT tenodesis. ASSA, however, does remove less cortical bone than does CIS and therefore produces a smaller stress riser in the proximal humerus. Further testing as to the integrity of ASSA is warranted to determine the integrity of the tenodesis with cyclical loading.

Для цитирования: Smuin Dallas M, Vannatta Emily, Ammerman Brittany, Stauch Christopher M, Lewis Gregory S, Dhawan Aman. Increased Load to Failure in Biceps Tenodesis With All-Suture Suture Anchor Compared With Interference Screw: A Cadaveric Biomechanical Study.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2021. DOI: 10.1016/j.arthro.2021.03.085.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.