One-year outcomes following repair of thoracoabdominal aneurysms with the multilayer flow modulator: report from the STRATO trial.

Vaislic Claude D, Fabiani Jean Noël, Chocron Sidney, Robin Jacques, Costache Victor S, Villemot Jean-Pierre, Alsac Jean Marc, Leprince Pascal N, Unterseeh Thierry, Portocarrero Eric, Glock Yves, Rousseau Hervé, STRATO Investigators Group

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate endovascular repair of type II and III thoracoabdominal aortic aneurysms (TAAA) using the Multilayer Flow Modulator (MFM) in patients with contraindications for open surgery and fenestrated stent-grafts.

METHODS

In this prospective, multicenter, nonrandomized trial (EudraCT registration: 2009-013678-42; ClinicalTrials.gov identifier NCT01756911), 23 patients (19 men; mean age 75.8 years) with Crawford type II (43.5%) and III (56.5%) TAAA (mean diameter 6.5 cm) were treated with the MFM between April 2010 and February 2011. The primary efficacy outcome measure was stable aneurysm thrombosis with associated branch vessel patency at 12 months; the primary safety endpoint was 30-day and 12-month all-cause mortality.

RESULTS

The rate of technical success was 100%. In 20 patients with computed tomography scans at 12 months, the primary efficacy outcome was met in 15 patients. The rate of primary patency of covered branch vessels was 96% (53/55); 1 patient with 2 occluded visceral branches underwent successful surgical reintervention. Endoleaks were identified in 5 patients (3 attachment site and 2 at device overlap), 4 of whom underwent reintervention (3 additional MFMs and 1 stent-graft implanted). At 12 months, aneurysm diameter was stable in 18 of 20 patients; the mean ratio of residual aneurysm flow volume to total volume had decreased by 28.9%, and the mean ratio of thrombus volume to total lumen volume had increased by 21.3% (n=17). There were no cases of device migration, loss of device integrity, spinal cord ischemia, or aneurysm rupture.

CONCLUSION

At 1 year, endovascular repair with the MFM appears to be safe and effective while successfully maintaining branch vessel patency. Follow-up is ongoing.

Для цитирования: Vaislic Claude D, Fabiani Jean Noël, Chocron Sidney, Robin Jacques, Costache Victor S, Villemot Jean-Pierre, Alsac Jean Marc, Leprince Pascal N, Unterseeh Thierry, Portocarrero Eric, Glock Yves, Rousseau Hervé, STRATO Investigators Group. One-year outcomes following repair of thoracoabdominal aneurysms with the multilayer flow modulator: report from the STRATO trial.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2014. DOI: 10.1583/13-4553r.1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.