[Investigation of the Usefulness of Monte Carlo Simulation in Electron Beam Therapy Using Body Surface Lead Cutout].
Аннотация
PURPOSE: The purpose of this study was to understand the PDD and OAR during electron beam therapy using lead cutout on the body surface.
METHODS
The Monte Carlo code PHITS version 3.24 was used to simulate PDD and OAR. The simulation results were compared with actual measurements using a silicon diode detector to evaluate the validity of the simulation results.
RESULTS
The simulated PDD and OAR parameters of the linac agreed with the measured values within 2 mm. When the lead cutout on the body surface was used, all parameters except for R100 agreed with the measured values within 2 mm. The cutout sizes of the broad-beam square irradiation fields were 3 cm for 6 MeV, 5 cm for 12 MeV, and 8 cm for 18 MeV when the lead cutout on the body surface was used.
CONCLUSION
The Monte Carlo simulation was useful for understanding the PDD and OAR of the lead cutout irradiation fields, which are difficult to measure.