Pre-analytical phase: The automated ProTube device supports quality assurance in the phlebotomy process.

Piva Elisa, Tosato Francesca, Plebani Mario

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Most errors in laboratory medicine occur in the pre-analytical phase of the total testing process. Phlebotomy, a crucial step in the pre-analytical phase influencing laboratory results and patient outcome, calls for quality assurance procedures and automation in order to prevent errors and ensure patient safety.

METHODS

We compared the performance of a new small, automated device, the ProTube Inpeco, designed for use in phlebotomy with a complete traceability of the process, with a centralized automated system, BC ROBO.

RESULTS

ProTube was used for 15,010 patients undergoing phlebotomy with 48,776 tubes being labeled. The mean time and standard deviation (SD) for blood sampling was 3:03 (min:sec; SD ± 1:24) when using ProTube, against 5:40 (min:sec; SD ± 1:57) when using BC ROBO. The mean number of patients per hour managed at each phlebotomy point was 16 ± 3 with ProTube, and 10 ± 2 with BC ROBO. No tubes were labeled erroneously or incorrectly, even if process failure occurred in 2.8% of cases when ProTube was used.

CONCLUSIONS

Thanks to its cutting edge technology, the ProTube has many advantages over BC ROBO, above all in verifying patient identity, and in allowing a reduction in both identification error and tube mislabeling.

Для цитирования: Piva Elisa, Tosato Francesca, Plebani Mario. Pre-analytical phase: The automated ProTube device supports quality assurance in the phlebotomy process.. Clinica chimica acta; international journal of clinical chemistry. 2015. DOI: 10.1016/j.cca.2015.10.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.