MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/14992027.2023.2169200

Медицинская библиотека · International journal of audiology · 2023

Predicting children's real-ear-to-coupler differences based on tympanometric data.

McCreery Ryan W, Crukley Jeffery, Grindle Anastasia, Merchant Gabrielle R, Walker Elizabeth

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

OBJECTIVE

Paediatric hearing-aid verification relies on measures of output obtained from the ear canal or in a coupler with the child's real-ear-to-coupler difference (RECD). Measured RECD cannot always be completed in children, leading to fitting inaccuracies. Audiologists often have tympanometry data that characterises the child's ear-canal acoustics. The goal of this study was to determine if tympanometry can be used to improve predictions of measured RECD.

DESIGN

A retrospective analysis of RECD and admittance, tympanometric peak pressure, and equivalent ear-canal volume from 226 Hz tympanometry collected as part of a longitudinal study of children with hearing loss were modelled with Bayesian hierarchical regression.

STUDY SAMPLE: Two-hundred sixty-six children with mild-to-severe hearing loss contributed data.

RESULTS

Age-based average RECD models were within 3 dB of measured RECD values in 54% of cases with normal middle ear status and 50.6% of cases with abnormal middle ear status. Immittance-predicted RECD were within 3 dB in 69.6% of cases with normal middle ear status and 74.4% of cases with abnormal middle ear status.

CONCLUSION

Immittance-predicted RECD was more accurate than age-based average RECD, particularly in children with abnormal middle ear status. The findings suggest that 226 Hz tympanometry could be used clinically to improve predictions of measured RECD when it cannot be measured.

Для цитирования: McCreery Ryan W, Crukley Jeffery, Grindle Anastasia, Merchant Gabrielle R, Walker Elizabeth. Predicting children's real-ear-to-coupler differences based on tympanometric data.. International journal of audiology. 2023. DOI: 10.1080/14992027.2023.2169200.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.