DOI 10.1016/j.redar.2016.04.008
Assessment of TruviewPCD videolaryngoscope for intubation in children.
Аннотация
OBJECTIVE
The aim of this observational prospective study was to evaluate the usefulness of TruviewPCD for tracheal intubation in clinical practice, and to provide data for future studies.
MATERIAL AND METHOD: A study was conducted on 86 consecutive children undergoing ear, nose and throat (ENT) or paediatric procedures under general anaesthesia with tracheal intubation. Children with two or more difficult airway criteria were excluded. A descriptive statistical analysis was performed.
RESULTS
Eighty-three patients were successfully intubated with TruviewPCD. Demographic data: Age 4.9 (2.8) years, weight 19.5 (7.7)kg. Seventy-nine children needed one attempt and four required two attempts at intubation. Time for glottis view and tracheal intubation was 10.8 (5.6) and 30 [27.9-37] seconds, respectively. Eighty-one patients were classified as easy or very easy to intubate, and only two cases were considered difficult. No significant complications were registered.
CONCLUSIONS
TruviewPCD is a good device for paediatric airway management. It would be interesting to have an intermediate blade between size 1 and 2, as the difference between both is too wide.