MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/eja.0000000000000367

Медицинская библиотека · European journal of anaesthesiology · 2016

McGRATH MAC video laryngoscope for insertion of a transoesophageal echocardiography probe: A randomised controlled trial.

Ishida Takashi, Kiuchi Chiaki, Sekiguchi Takemi, Tsujimoto Takatoshi, Kawamata Mikito

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Transoesophageal echocardiography (TOE) probe insertion in anaesthetised patients can cause pharyngeal and oesophageal injuries. Kim et al. have shown that insertion assisted by a Macintosh laryngoscope can reduce such complications but it may sometimes be difficult to observe the passage of a TOE probe. The McGRATH MAC (McGRATH) has been shown to provide a better view of the glottis, piriform fossa and oesophageal inlet during tracheal intubation than the Macintosh.

OBJECTIVE

We hypothesised that the McGRATH provided better visualisation of the oesophageal inlet and was useful as an aid to TOE probe placement, possibly reducing the incidence of pharyngeal injury related to insertion, compared with the Macintosh.

DESIGN

A randomised controlled trial.

SETTING

The study was conducted in a university hospital from February to December 2014.

PATIENTS: One hundred patients undergoing elective surgery under intraoperative TOE monitoring were randomised to either a Macintosh group or a McGRATH group.

INTERVENTIONS: Macintosh and McGRATH were used to visualise the passage of the TOE probe and guide its insertion.

MAIN OUTCOME MEASURES: Visibility of the oesophageal inlet, the number of TOE insertion attempts and incidence of pharyngeal mucosal injury after the TOE probe had been removed were assessed.

RESULTS

The percentage of patients in whom the oesophageal inlet was visible was higher in the McGRATH group (88%) than in the Macintosh group (41%) (P < 0.01). The number of TOE probe insertion attempts was significantly smaller in the McGRATH group than in the Macintosh group (P = 0.039). The incidence of pharyngeal mucosal injury was significantly smaller in the McGRATH group (4%) than in the Macintosh group (16%; P = 0.042).

CONCLUSION

The McGRATH provided a better view of the oesophageal inlet and was useful as an aid to TOE probe placement, possibly reducing the incidence of pharyngeal injury related to its insertion.

TRIAL REGISTRATION: University Hospital Medical Information Network in Japan (UMIN) 000012970.

Для цитирования: Ishida Takashi, Kiuchi Chiaki, Sekiguchi Takemi, Tsujimoto Takatoshi, Kawamata Mikito. McGRATH MAC video laryngoscope for insertion of a transoesophageal echocardiography probe: A randomised controlled trial.. European journal of anaesthesiology. 2016. DOI: 10.1097/eja.0000000000000367.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.