MedElevator Войти в исследования
← Поиск публикаций
BMC genetics · 2018

High resolution melting curve analysis enables rapid and reliable detection of G6PD variants in heterozygous females.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Islam Md Tarikul, Sarker Suprovath Kumar, Talukder Shezote, Bhuyan Golam Sarower, Rahat Asifuzzaman, Islam Nafisa Nawal, Mahmud Hasan, Hossain Mohammad Amir, Muraduzzaman A K M, Rahman Jakia, Qadri Syeda Kashfi, Shahidullah Mohammod, Mannan Mohammad Abdul, Tahura Sarabon, Hussain Manzoor, Saha Narayan, Akhter Shahida, Nahar Nazmun, Begum Firoza, Shirin Tahmina, Akhteruzzaman Sharif, Qadri Syed Saleheen, Qadri Firdausi, Mannoor Kaiissar

Аннотация

BACKGROUND: Like glucose-6-phosphate dehydrogenase (G6PD) deficient hemizygous males and homozygous females, heterozygous females could also manifest hemolytic crisis, neonatal hyperbilirubinemia or kernicterus upon exposure to oxidative stress induced by certain foods such as fava beans, drugs or infections. Although hemizygous males and homozygous females are easily detected by conventional G6PD enzyme assay method, the heterozygous state could be missed by the conventional methods as the mosaic population of both normal and deficient RBCs circulates in the blood. Thus the present study aimed to apply high resolution melting (HRM) curve analysis approach to see whether HRM could be used as a supplemental approach to increase the chance of detection of G6PD heterozygosity.

RESULTS: Sixty-three clinically suspected females were evaluated for G6PD status using both enzyme assay and HRM analysis. Four out of sixty-three participants came out as G6PD deficient by the enzyme assay method, whereas HRM approach could identify nine participants with G6PD variants, one homozygous and eight heterozygous. Although only three out of eight heterozygous samples had G6PD enzyme deficiency, the HRM-based heterozygous G6PD variants detection for the rest of the samples with normal G6PD enzyme activities could have significance because their newborns might fall victim to serious consequences under certain oxidative stress.

CONCLUSIONS: In addition to the G6PD enzyme assay, HRM curve analysis could be useful as a supplemental approach for detection of G6PD heterozygosity.