A clinical evaluation of the ProNOVA XR polymer-free sirolimus eluting coronary stent system in the treatment of patients with de novo coronary artery lesions (EURONOVA XR I study).
Аннотация
AIMS: Evaluation of safety and efficacy of ProNOVA XR, a new generation of polymer-free sirolimus eluting stents (SES), utilizing a pharmaceutical excipient for timed release of sirolimus from the XR platform.
METHODS AND
RESULTS
Safety and efficacy of ProNOVA XR coronary stent system was examined in EURONOVA prospective, single arm, multi-center registry of 50 patients with de novo native coronary lesions up to 28 mm in length in arteries between 2.25 and 4 mm. At 6-month, in-stent late lumen loss by QCA was 0.45 ± 0.41 mm and in-stent neointimal volume obstruction in the IVUS sub-study was 14 ± 11%. One-year clinical follow-up revealed a favorable safety profile, with 2% of in-hospital MACE and 6.4% of MACE from hospital discharge up to 12 months (including 1 cardiac death >30 days after stent implantation and 2 TLRs). According to the ARC definition, there was no definite or probable stent thrombosis and 1 possible stent thrombosis (2%) up to 12 months of clinical follow-up.
CONCLUSIONS
In this preliminary evaluation, ProNOVA XR polymer-free sirolimus eluting stent system appeared safe with an early promise of adequate effectiveness in the treatment of de novo coronary lesions in up to 12 months of clinical, angiographic and IVUS follow-up.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.