Rescue of Failed Aortic Repair with Fenestrated Endovascular Device.
Аннотация
INTRODUCTION
The incidence of failed endovascular (EVAR) and open repair (OR) is increasing. Redo aortic repair is required in 10% of patients. Extension of the proximal sealing zone above the visceral arteries to adequate, healthier thoracic aorta using a fenestrated graft (FEVAR) can rescue a failing repair. A custom-made device can treat proximal type 1a endoleaks or proximal dilatation post endovascular or open repair, respectively. The aim of this investigation was to present a single-centre experience with FEVAR for patients with a failing aortic repair.
METHODS
A prospectively maintained database of FEVAR patients treated with a ZenithⓇ Fenestrated endovascular (ZFEN) device (Cook Medical LLC, Bloomington, Indiana, USA) was interrogated for individuals who had the device implanted as a rescue therapy after prior endovascular (EVAR) or open repair (OR). Statistical analysis was performed with SPSS v 25 software.
RESULTS
Between January 1, 2011 and March 31, 2019, 17 ZFEN devices were implanted. 10 patients had a type 1a endoleak from a prior EVAR and 7 patients had proximal disease progression after prior OR. There were 12 males and 5 females, median age of 75 (interquartile range, IQR 7). 76.4% (n = 13) of patients had an American Society of Anaesthesiologists (ASA) grade of 3. Primary technical success was 70.5% (n = 12). Of the remainder, 4 cases (24%) had a type III endoleak at completion angiogram; of which, 2 patients (12%) required re-intervention within 30 days. One further case (6%) had primary assisted technical success as stenting of a flow limiting dissection flap in an iliac vessel was required. Peri-operative rate of deployment related complications and systemic complications were 5.8% (n = 1) and 35% (n = 6), respectively. Median length of hospital stay was 11 days (IQR 11). There was no mortality within the study follow up. Overall 30-day re-intervention rate was 23.5%. Overall survival was 92% at one year.
CONCLUSION
FEVAR is a safe but technically challenging option for rescue of failing aortic repairs. These are a high-risk group of patients and this is reflected in the high post-operative morbidity rate. Technical success was high and 30-day mortality was low.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.