Efficacy of two anatomically preformed supraglottic devices, the BlockBuster Laryngeal mask airway and the Ambu AuraGain, for ventilation in anesthetised paralyzed pediatric patients: a randomized trial.

Arumugasamy Vignesh, Agrawal Nidhi, Gupta Anju

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The BlockBuster Laryngeal Mask Airway (LMA) provides a higher oropharyngeal seal pressure (OSP) than Ambu Aura Gain (AAG) in children at an intracuff pressure of 60cmH2O. Lower intracuff pressures prevent laryngopharyngeal morbidity; therefore, we compared their OSP at 40cmH2O.

METHODS

Seventy children scheduled to undergo elective surgery under general anesthesia were randomly assigned to the AAG (Group A) or BlockBuster LMA (Group B) groups. After device placement, the cuff was inflated to 40cmH2O pressure. Ease of insertion, OSP, number of attempts, insertion time, glottic alignment, peak airway pressure (PAP) to OSP difference (OSP-PAP), and leak fraction were recorded.

RESULTS

Mean OSP was higher in Group B at 5 min (29.7 ± 1.4 vs. 24.9 ± 2.1cmH2O; p < 0.0001) and 30 min (29.6 ± 1.3 vs. 24.1 ± 2.6; p < 0.0001). OSP-PAP values were significantly higher in Group B at 5 and 30 min (p < 0.0001). Anatomical alignment was significantly better in Group B (88.6% vs. 57.1%; p = 0.006). The mean times to achieve effective ventilation, ease of device insertion, and gastric catheter insertion were comparable (p = 1.00). Leakage percentages were similar at 5 min (p = 0.11) and 30 min (p = 0.06).            .

CONCLUSION

The BlockBuster LMA provided higher OSP, lower PAP-OSP, and better glottic alignment. It may be preferred in children scheduled for elective surgery.

CLINICAL TRIAL NUMBER AND REGISTRY URL: CTRI/2024/06/068284 [Registered on: 04/06/2024] Trial Registered Prospectively.

Для цитирования: Arumugasamy Vignesh, Agrawal Nidhi, Gupta Anju. Efficacy of two anatomically preformed supraglottic devices, the BlockBuster Laryngeal mask airway and the Ambu AuraGain, for ventilation in anesthetised paralyzed pediatric patients: a randomized trial.. Expert review of medical devices. 2026. DOI: 10.1080/17434440.2026.2695251.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.