Diagnostic accuracy of a hand-held ultrasound scanner in routine patients referred for echocardiography.
Аннотация
BACKGROUND
The aim of this study was to investigate the imaging capabilities of recent hand-held ultrasound scanners.
METHODS
Three hundred forty-nine patients were scanned with hand-held ultrasound (HAND) and high-end echocardiography (HIGH). Segmental endocardial border delineation was scored (2 = good, 1 = poor, 0 = invisible) to describe image quality. Assessments of left ventricular (LV) dimensions, regional and global LV function, and grades of valve disease were compared.
RESULTS
The mean endocardial visibility grades were 1.6 ± 0.5 with HAND and 1.7 ± 0.4 with HIGH (P < .01). Regional wall motion was scored very similarly (κ = 0.73, P < .01). Ejection fraction assessment (bias = 1.8%, 1.96 × SD = 8.3%) and LV measurements (r = 0.99, P < .01; interventricular septum: bias = 0.91 mm, 1.96 × SD = 2.1 mm; LV end-diastolic diameter: bias = 0.5 mm, 1.96 × SD = 4.1 mm; LV posterior wall: bias = 0.61 mm, 1.96 × SD = 2.4 mm) showed negligible deviations. No pericardial effusion or valve stenosis was missed. Regurgitations missed by HAND were all graded "minimal" on HIGH. Regurgitations were mildly overestimated by HAND. Overall concordance for detection of regurgitations was very good (κ = 0.9, P < .01).
CONCLUSIONS
Handheld echocardiography was feasible and missed no relevant findings. Given the future implementation of spectral Doppler capabilities, this handheld scanner can safely be used in clinical routine.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.