DOI 10.1007/s10006-022-01050-5

Медицинская библиотека · Oral and maxillofacial surgery · 2022

Accuracy of low-cost alternative facial scanners: a prospective cohort study.

Bartella Alexander K, Laser Josefine, Kamal Mohammad, Krause Matthias, Neuhaus Michael, Pausch Niels C, Sander Anna K, Lethaus Bernd, Zimmerer Rüdiger

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Three-dimensional facial scans have recently begun to play an increasingly important role in the peri-therapeutic management of oral and maxillofacial and head and neck surgery cases. Face scan images can be generated by optical facial scanners utilizing line-laser, stereophotography, or structured light modalities, as well as from volumetric data: for example, from cone beam computed tomography (CBCT). This study aimed to evaluate whether two low-cost procedures for the creation of three-dimensional face scan images were capable of producing sufficiently accurate data sets for clinical analysis.

MATERIALS AND

METHODS

Fifty healthy volunteers were included in the study. Two test objects with defined dimensions (Lego bricks) were attached to the forehead and the left cheek of each volunteer. Facial anthropometric values (i.e., the distances between the medial canthi, the lateral canthi, the nasal alae, and the angles of the mouth) were first measured manually. Subsequently, face scans were performed with a smart device and manual photogrammetry and the values obtained were compared with the manually measured data sets.

RESULTS

The anthropometric distances deviated, on average, 2.17 mm from the manual measurements (smart device scanning deviation 3.01 mm, photogrammetry deviation 1.34 mm), with seven out of eight deviations being statistically significant. For the Lego brick, from a total of 32 angles, 19 values demonstrated a significant difference from the original 90° angles. The average deviation was 6.5° (smart device scanning deviation 10.1°, photogrammetry deviation 2.8°).

CONCLUSION

Manual photogrammetry demonstrated greater accuracy when creating three-dimensional face scan images; however, smart devices are more user-friendly. Dental professionals should monitor camera and smart device technical improvements carefully when choosing and adequate technique for 3D scanning.

Для цитирования: Bartella Alexander K, Laser Josefine, Kamal Mohammad, Krause Matthias, Neuhaus Michael, Pausch Niels C, Sander Anna K, Lethaus Bernd, Zimmerer Rüdiger. Accuracy of low-cost alternative facial scanners: a prospective cohort study.. Oral and maxillofacial surgery. 2022. DOI: 10.1007/s10006-022-01050-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.