Deep learning reconstruction for improving image quality of pediatric abdomen MRI using a 3D T1 fast spoiled gradient echo acquisition.

Zucker Evan J, Milshteyn Eugene, Machado-Rivas Fedel A, Tsai Leo L, Roberts Nathan T, Guidon Arnaud, Gee Michael S, Victoria Teresa

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Deep learning (DL) reconstructions have shown utility for improving image quality of abdominal MRI in adult patients, but a paucity of literature exists in children.

OBJECTIVE

To compare image quality between three-dimensional fast spoiled gradient echo (SPGR) abdominal MRI acquisitions reconstructed conventionally and using a prototype method based on a commercial DL algorithm in a pediatric cohort.

MATERIALS AND

METHODS

Pediatric patients (age < 18 years) who underwent abdominal MRI from 10/2023-3/2024 including gadolinium-enhanced accelerated 3D SPGR 2-point Dixon acquisitions (LAVA-Flex, GE HealthCare) were identified. Images were retrospectively generated using a prototype reconstruction method leveraging a commercial deep learning algorithm (AIR™ Recon DL, GE HealthCare) with the 75% noise reduction setting. For each case/reconstruction, three radiologists independently scored DL and non-DL image quality (overall and of selected structures) on a 5-point Likert scale (1-nondiagnostic, 5-excellent) and indicated reconstruction preference. The signal-to-noise ratio (SNR) and mean number of edges (inverse correlate of image sharpness) were also quantified. Image quality metrics and preferences were compared using Wilcoxon signed-rank, Fisher exact, and paired t-tests. Interobserver agreement was evaluated with the Kendall rank correlation coefficient (W).

RESULTS

The final cohort consisted of 38 patients with mean ± standard deviation age of 8.6 ± 5.7 years, 23 males. Mean image quality scores for evaluated structures ranged from 3.8 ± 1.1 to 4.6 ± 0.6 in the DL group, compared to 3.1 ± 1.1 to 3.9 ± 0.6 in the non-DL group (all P < 0.001). All radiologists preferred DL in most cases (32-37/38, P < 0.001). There were a 2.3-fold increase in SNR and a 3.9% reduction in the mean number of edges in DL compared to non-DL images (both P < 0.001). In all scored anatomic structures except the spine and non-DL adrenals, interobserver agreement was moderate to substantial (W = 0.41-0.74, all P < 0.01).

CONCLUSION

In a broad spectrum of pediatric patients undergoing contrast-enhanced Dixon abdominal MRI acquisitions, the prototype deep learning reconstruction is generally preferred to conventional methods with improved image quality across a wide range of structures.

Для цитирования: Zucker Evan J, Milshteyn Eugene, Machado-Rivas Fedel A, Tsai Leo L, Roberts Nathan T, Guidon Arnaud, Gee Michael S, Victoria Teresa. Deep learning reconstruction for improving image quality of pediatric abdomen MRI using a 3D T1 fast spoiled gradient echo acquisition.. Pediatric radiology. 2025. DOI: 10.1007/s00247-025-06313-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.