DOI 10.1259/bjr.20140862

Медицинская библиотека · The British journal of radiology · 2015

Dosimetric effects of roll rotational setup errors on lung stereotactic ablative radiotherapy using volumetric modulated arc therapy.

Lee Jaegi, Kim Jung-In, Ye Sung-Joon, Kim Hak Jae, Carlson Joel, Park Jong Min

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the dosimetric effects of roll-rotational setup errors of stereotactic ablative radiotherapy (SABR) for lung cancer using volumetric modulated arc therapy (VMAT).

METHODS

A total of 23 lung SABR cases were evaluated retrospectively. Each of the planning CT images was intentionally rotated by ±1°, ±2° and ±3°. After that, to simulate the translational couch correction, rotated CT images were moved along the x, y and z axis to match the centroid of the target volume in the rotated CT images with that in the original CT images. The differences in D95% and V100% of the target volume, D0.35cc of spinal cord, D0.35cc and D5cc of oesophagus and V20Gy of lung between the original and the rotated CT images were calculated.

RESULTS

The average differences in D95% and V100% of target volume, D0.35cc of spinal cord, D0.35cc and D5cc of oesophagus and V20Gy of lung were -0.3% ± 0.4% and -0.7% ± 2.4%, 1.6 ± 27.9 cGy, -1.6 ± 37.6 cGy, 15.9 ± 25.3 cGy and 0.0% ± 0.1%, respectively. The dosimetric changes in organs at risk (OARs) near the target volume were sometimes considerable due to roll-rotational setup errors, despite the translational correction, and those were patient specific.

CONCLUSION

In the case of coplanar VMAT for lung SABR, dosimetric changes to the target volume due to roll-rotational setup errors could be compensated by translational correction, whereas those to the OARs could not in some cases.

ADVANCES IN KNOWLEDGE: Roll-rotational setup errors would increase the dose to OARs despite the translational correction.

Для цитирования: Lee Jaegi, Kim Jung-In, Ye Sung-Joon, Kim Hak Jae, Carlson Joel, Park Jong Min. Dosimetric effects of roll rotational setup errors on lung stereotactic ablative radiotherapy using volumetric modulated arc therapy.. The British journal of radiology. 2015. DOI: 10.1259/bjr.20140862.