Angiology · 2009
Angio-Seal use in patients with peripheral arterial disease (ASPIRE).
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Katzenschlager Reinhold, Tischler Reinhold, Kalchhauser Georg, Panny Michael, Hirschl Mirko
Аннотация
PURPOSE: To investigate the incidence of complications after the use of an arterial closure device (Angio-Seal) in patients with peripheral arterial disease.
METHODS: In 105 consecutive patients after transfemoral catheterization, the puncture site was closed using a closure device (Angio-Seal). Colour-flow-duplexsonography studies were conducted 1 to 4 days before, within 3 days after and 3 month after the intervention.
RESULTS: All patients had peripheral arterial disease, 34 had calcification at the puncture site. Detection of calcification did not prevent device deployment. Complications (2 minor bleedings, 1 pseudoaneurysm) were not associated with high risk groups (these were: 69 antegrade punctures, 22 obese and 32 hypertensive patients). Three-month postinterventional diameter and blood velocity changes were <1%.
CONCLUSIONS: Patients with peripheral arterial disease in the region of the puncture site and patients at higher complication risk can safely and effectively be closed with an Angio-Seal device. At the puncture site, no lumen change can be observed 3 months postinterventional.
METHODS: In 105 consecutive patients after transfemoral catheterization, the puncture site was closed using a closure device (Angio-Seal). Colour-flow-duplexsonography studies were conducted 1 to 4 days before, within 3 days after and 3 month after the intervention.
RESULTS: All patients had peripheral arterial disease, 34 had calcification at the puncture site. Detection of calcification did not prevent device deployment. Complications (2 minor bleedings, 1 pseudoaneurysm) were not associated with high risk groups (these were: 69 antegrade punctures, 22 obese and 32 hypertensive patients). Three-month postinterventional diameter and blood velocity changes were <1%.
CONCLUSIONS: Patients with peripheral arterial disease in the region of the puncture site and patients at higher complication risk can safely and effectively be closed with an Angio-Seal device. At the puncture site, no lumen change can be observed 3 months postinterventional.
Предметные рубрики MeSH
Aneurysm, FalseBlood Flow VelocityCalcinosisCatheterization, PeripheralEquipment DesignFemaleFemoral ArteryHemorrhageHemostatic TechniquesHumansHypertensionLaser-Doppler FlowmetryMaleObesityPeripheral Vascular DiseasesPuncturesRadiographyRegional Blood FlowRisk AssessmentTime FactorsTreatment OutcomeUltrasonography, Doppler, Color