Enhancement of a prosthetic knee with a microprocessor-controlled gait phase switch reduces falls and improves balance confidence and gait speed in community ambulators with unilateral transfemoral amputation.
Аннотация
BACKGROUND
Despite the evidence for improved safety and function of microprocessor stance and swing-controlled prosthetic knees, non-microprocessor-controlled prosthetic knees are still standard of care for persons with transfemoral amputations in most countries. Limited feature microprocessor-control enhancement of such knees could stand to significantly improve patient outcomes.
OBJECTIVES
To evaluate gait speed, balance, and fall reduction benefits of the new 3E80 default stance hydraulic knee compared to standard non-microprocessor-controlled prosthetic knees.
STUDY
DESIGN
Comparative within-subject clinical study.
METHODS
A total of 13 young, high-functioning community ambulators with a transfemoral amputation underwent assessment of performance-based (e.g. 2-min walk test, timed ramp/stair tests) and self-reported (e.g. falls, Activities-Specific Balance Confidence scale, Prosthesis Evaluation Questionnaire question #1, Satisfaction with the Prosthesis) outcome measures for their non-microprocessor-controlled prosthetic knees and again after 8 weeks of accommodation to the 3E80 microprocessor-enhanced knee.
RESULTS
Self-reported falls significantly declined 77% ( p = .04), Activities-Specific Balance Confidence scores improved 12 points ( p = .005), 2-min walk test walking distance increased 20 m on level ( p = .01) and uneven ( p = .045) terrain, and patient satisfaction significantly improved ( p < .01) when using the 3E80 knee. Slope and stair ambulation performance did not differ between knee conditions.
CONCLUSION
The 3E80 knee reduced self-reported fall incidents and improved balance confidence. Walking performance on both level and uneven terrains also improved compared to non-microprocessor-controlled prosthetic knees. Subjects' satisfaction was significantly higher than with their previous non-microprocessor-controlled prosthetic knees. The 3E80 may be considered a prosthetic option for improving gait performance, balance confidence, and safety in highly active amputees. Clinical relevance This study compared performance-based and self-reported outcome measures when using non-microprocessor and a new microprocessor-enhanced, default stance rotary hydraulic knee. The results inform rehabilitation professionals about the functional benefits of a limited-feature, microprocessor-enhanced hydraulic prosthetic knee over standard non-microprocessor-controlled prosthetic knees.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.