Predictors of increase in pacing threshold after transcatheter pacing system implantation due to micro-dislodgement.

Higuchi Motoaki, Shinoda Yasutoshi, Hasegawa Tomoaki, Ishibashi Mayu, Yamada Norihiro, Chiba Yoshiro, Ohira Koji, Murata Minoru, Aonuma Kazutaka

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Achieving a favorable pacing threshold with a Micra transcatheter pacing system (Micra-TPS) is needed to reduce battery depletion. In some cases, the threshold increases shortly after the device is implanted, and a higher pacing threshold may be required. This study aims to identify the causes and predictors of the increase in pacing threshold observed shortly after Micra-TPS implantation.

METHODS

The study included 64 consecutive patients who underwent Micra-TPS implantation between 2017 and 2020. The patients were divided into two groups depending on their pacing threshold: the increased pacing threshold (IPT) group (threshold increased by ≥0.5 V/0.24 ms within 1 month of implantation) and the stable pacing threshold (SPT) group.

RESULTS

Excluding four patients who could not be followed up, of the 60 remaining patients, nine (15%) were in the IPT group and 51 (85%) were in the SPT group. The IPT group had significantly lower implant impedance values and higher implant thresholds than the SPT group (582 ± 59 vs 755 ± 167 Ω [P < .001] and 1.29 ± 0.87 vs 0.71 ± 0.40 V/0.24 ms [P = .014]). Implant impedance and threshold may serve as predictors of a threshold increase after implantation (area under the curve: 0.737-0.943 and 0.586-0.926, respectively).

CONCLUSIONS

An IPT was noted shortly after Micra-TPS implantation owing to micro-dislodgement because of insufficient anchoring of the device to the myocardium. Impedance >660 Ω and threshold <1.0 V/0.24 ms may predict an increase in pacing threshold.

Для цитирования: Higuchi Motoaki, Shinoda Yasutoshi, Hasegawa Tomoaki, Ishibashi Mayu, Yamada Norihiro, Chiba Yoshiro, Ohira Koji, Murata Minoru, Aonuma Kazutaka. Predictors of increase in pacing threshold after transcatheter pacing system implantation due to micro-dislodgement.. Pacing and clinical electrophysiology : PACE. 2020. DOI: 10.1111/pace.14080.