[Verification of Dose Distribution in Cervical Cancer Brachytherapy Using Metal and Plastic Applicators].
Аннотация
PURPOSE: To validate the point-A dose and dose distribution of metal and resin applicators in comparison with those of TG-43U1.
METHODS
The metal and resin applicators consisting of tandem and ovoid were modeled by the egs_brachy. The doses to point A and dose distributions considering each applicator were calculated and compared to those of TG-43U1.
RESULTS
The dose to point A considering the metal applicator was 3.2% lower than that of TG-43U1, but there was no difference in the dose to point A considering the resin applicator. The dose distribution considering the metal applicator was lower than that of TG-43U1 at all calculation points, but there was no difference in the dose distribution considering the resin applicator at almost all calculation points.
CONCLUSION
In this study, the dose distribution considering the metal applicator was lower than that of TG-43U1 at all calculation points, but there was no difference in the dose distribution considering the resin applicator at almost all calculation points. Therefore, TG-43U1 can accurately calculate the dose distribution when changing from the metal applicator to the resin applicator.