[Verification of Dose Distribution in Cervical Cancer Brachytherapy Using Metal and Plastic Applicators].

Tomita Fumihiro, Fuse Hiraku, Fujisaki Tatsuya, Yasue Kenji, Miyakawa Sin, Ikoma Hideaki, Yoshizawa Tomoyuki, Oyama Katsuhiko, Oyama Satoshi, Okumura Toshiyuki, Tamaki Yoshio

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To validate the point-A dose and dose distribution of metal and resin applicators in comparison with those of TG-43U1.

METHODS

The metal and resin applicators consisting of tandem and ovoid were modeled by the egs_brachy. The doses to point A and dose distributions considering each applicator were calculated and compared to those of TG-43U1.

RESULTS

The dose to point A considering the metal applicator was 3.2% lower than that of TG-43U1, but there was no difference in the dose to point A considering the resin applicator. The dose distribution considering the metal applicator was lower than that of TG-43U1 at all calculation points, but there was no difference in the dose distribution considering the resin applicator at almost all calculation points.

CONCLUSION

In this study, the dose distribution considering the metal applicator was lower than that of TG-43U1 at all calculation points, but there was no difference in the dose distribution considering the resin applicator at almost all calculation points. Therefore, TG-43U1 can accurately calculate the dose distribution when changing from the metal applicator to the resin applicator.

Для цитирования: Tomita Fumihiro, Fuse Hiraku, Fujisaki Tatsuya, Yasue Kenji, Miyakawa Sin, Ikoma Hideaki, Yoshizawa Tomoyuki, Oyama Katsuhiko, Oyama Satoshi, Okumura Toshiyuki, Tamaki Yoshio. [Verification of Dose Distribution in Cervical Cancer Brachytherapy Using Metal and Plastic Applicators].. Nihon Hoshasen Gijutsu Gakkai zasshi. 2023. DOI: 10.6009/jjrt.2023-1375.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.