DOI 10.3233/xst-230073

Медицинская библиотека · Journal of X-ray science and technology · 2023

Dosimetric properties of PASSAG polymer gel dosimeter in electron beam radiotherapy using magnetic resonance imaging.

Zhang Tiancheng, Almajidi Yasir Q, Awad Sameer A, Alhachami Firas Rahi, Gatea Maher Abdulfadhil, Kadhum Wesam R

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Several physical factors such as photon beam energy, electron beam energy, and dose rate may affect the dosimetric properties of polymer gel dosimeters. The photon beam energy and dose rate dependence of PASSAG gel dosimeter were previously evaluated.

OBJECTIVE

This study aims to assess the dosimetric properties of the optimized PASSAG gel samples in various electron beam energies.

METHODS

The optimized PASSAG gel samples are first fabricated and irradiated to various electron energies (5, 7, 10 and 12 MeV). Then, the response (R2) and sensitivity of gel samples are analyzed by magnetic resonance imaging technique at a dose range of 0 to 10 Gy, scanning room temperature range of 15 to 22 °C, and post-irradiation time range of 1 to 30 days.

RESULTS

The R2-dose response and sensitivity of gel samples do not change under the evaluated electron beam energies (the differences are less than 5%). Furthermore, a dose resolution range of 11 to 38 cGy is obtained for the gel samples irradiated to different electron beam energies. Moreover, the findings show that the R2-dose response and sensitivity dependence of gel samples on electron beam energy varies over different scanning room temperatures and post-irradiation times.

CONCLUSION

The dosimetric assessment of the optimized PASSAG gel samples provides the promising data for this dosimeter during electron beam radiotherapy.

Для цитирования: Zhang Tiancheng, Almajidi Yasir Q, Awad Sameer A, Alhachami Firas Rahi, Gatea Maher Abdulfadhil, Kadhum Wesam R. Dosimetric properties of PASSAG polymer gel dosimeter in electron beam radiotherapy using magnetic resonance imaging.. Journal of X-ray science and technology. 2023. DOI: 10.3233/xst-230073.