OBJECTIVE

Due to evidence for proton beam therapy (PBT) in pediatric central nervous system (CNS) tumors, compact proton therapy systems became commercially available to allow better integration in a hospital setting. However, these systems have a non-zero magnetic field at the level of the patient. Often, these patients have a cerebrospinal fluid shunt, and most of them are adjustable through a magnet. Whether the induced magnetic fields could interfere with adjustable shunts is unknown.

METHODS

In the first five CNS tumor patients with adjustable shunts who underwent PBT, the shunt setting was controlled before, during, and after treatment with PBT. Additionally, we used an ex vivo adjustable shunt to check if the settings could be altered by the magnetic field.

RESULTS

We did not observe unintentional changes in shunt settings in vivo during treatment. In ex vivo testing, the shunt settings were altered directly cranial to the exit window of PBT due to the magnetic field.

CONCLUSION

Although we did not observe any shunt setting alteration during PBT in this small cohort, caution is warranted. Given the lack of high-volume data, there should be a low threshold for checking the shunt setting at the end of PBT therapy or in a symptomatic patient.

Для цитирования: Dietvorst Sofie, Poels Kenneth, Van Beek Karen, Peeters Ronald, De Vloo Philippe. Adjustable shunts and proton therapy: a magnetic combination.. Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery. 2023. DOI: 10.1007/s00381-023-05984-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.