MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/j.2048-7940.1995.tb01629.x

Медицинская библиотека · Rehabilitation nursing : the official journal of the Association of Rehabilitation Nurses · 1995

Comparison study of home catheter cleaning methods.

Kurtz M J, Van Zandt K, Burns J L

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Three methods of cleaning urinary catheters for reuse at home by clients on intermittent catheterization programs were studied. Reused catheters were collected from clients, autoclaved, and then incubated in a culture of Escherichia coli broth. Three different isolates of E. coli were used at concentrations ranging from 4.8 x 10(5) to 1.0 x 10(8) colony-forming units per milliliter. The catheters were then rinsed with tap water for 1 minute and soaked in one of three cleaning solutions for 30 minutes. The three cleaning solutions studied were 0.6% hydrogen peroxide, bleach in a 1:4 solution with tap water, and betadine in a 1:2 solution with tap water. None of the cleaned catheters showed detectable growth for 48 hours after the cleaning procedure was performed.

Для цитирования: Kurtz M J, Van Zandt K, Burns J L. Comparison study of home catheter cleaning methods.. Rehabilitation nursing : the official journal of the Association of Rehabilitation Nurses. 1995. DOI: 10.1002/j.2048-7940.1995.tb01629.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.