MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1213/ane.0b013e31828610a0

Медицинская библиотека · Anesthesia and analgesia · 2013

Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Feiner John R, Rollins Mark D, Sall Jeffrey W, Eilers Helge, Au Paul, Bickler Philip E

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Carbon monoxide poisoning is a significant problem in most countries, and a reliable method of quick diagnosis would greatly improve patient care. Until the recent introduction of a multiwavelength "pulse CO-oximeter" (Masimo Rainbow SET(®) Radical-7), obtaining carboxyhemoglobin (COHb) levels in blood required blood sampling and laboratory analysis. In this study, we sought to determine whether hypoxemia, which can accompany carbon monoxide poisoning, interferes with the accurate detection of COHb.

METHODS

Twelve healthy, nonsmoking, adult volunteers were fitted with 2 standard pulse-oximeter finger probes and 2 Rainbow probes for COHb detection. A radial arterial catheter was placed for blood sampling during 3 interventions: (1) increasing hypoxemia in incremental steps with arterial oxygen saturations (SaO2) of 100% to 80%; (2) normoxia with incremental increases in %COHb to 12%; and (3) elevated COHb combined with hypoxemia with SaO2 of 100% to 80%. Pulse-oximeter (SpCO) readings were compared with simultaneous arterial blood values at the various increments of hypoxemia and carboxyhemoglobinemia (≈25 samples per subject). Pulse CO-oximeter performance was analyzed by calculating the mean bias (SpCO - %COHb), standard deviation of the bias (precision), and the root-mean-square error (A(rms)).

RESULTS

The Radical-7 accurately detected hypoxemia with both normal and elevated levels of COHb (bias mean ± SD: 0.44% ± 1.69% at %COHb <4%, and -0.29% ± 1.64% at %COHb ≥4%, P < 0.0001, and A(rms) 1.74% vs 1.67%). COHb was accurately detected during normoxia and moderate hypoxia (bias mean ± SD: -0.98 ± 2.6 at SaO2 ≥95%, and -0.7 ± 4.0 at SaO2 <95%, P = 0.60, and A(rms) 2.8% vs 4.0%), but when SaO2 decreased below approximately 85%, the pulse CO-oximeter always gave low signal quality errors and did not report SpCO values.

CONCLUSIONS

In healthy volunteers, the Radical-7 pulse CO-oximeter accurately detects hypoxemia with both low and elevated COHb levels, and accurately detects COHb, but only reads SpCO when SaO2 is more than approximately 85%.

Для цитирования: Feiner John R, Rollins Mark D, Sall Jeffrey W, Eilers Helge, Au Paul, Bickler Philip E. Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.. Anesthesia and analgesia. 2013. DOI: 10.1213/ane.0b013e31828610a0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.