Dosimetric correlation between exit dose maps and anatomical changes in head and neck cancer treatment on Halcyon-E: A gamma analysis approach.

Pereira da Silva Jayane Julia, Mendanha Ribeiro Caio Weber, Ludwig Jaime Luiz, Rocha Medrado Daniela, Melo da Silva Bruno Nunes, Alves Sarmento Ayane Cristine, de Moura Santos Edilmar

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Head and neck cancer (HNC) has a high incidence in Brazil. Although Image-Guided Radiation Therapy (IGRT) improves treatment precision, anatomical changes during the treatment course may compromise dose distribution and require replanning.

PURPOSE: This study investigated the correlation between exit dose fluence and anatomical volumetric variations during HNC radiotherapy. Subsequently, a decision support methodology was proposed and validated to flag potential dosimetric deviations and guide objective replanning strategies.

METHODS

A retrospective study was conducted with 11 patients, totaling 312 fractions treated on a Halcyon-E linear accelerator. Volumetric variations were derived from cone-beam computed tomography (CBCT) and exit dose fluence maps from each fraction were compared with the reference from the first treatment day using gamma analysis (γ_(1%/1 mm) to γ_(5%/5 mm); 10% threshold). Correlations between normalized volumetric variation (ΔV) and gamma indices were assessed using Kendall's tau coefficient (α = 0.05). A decision support methodology for replanning (γ_(1%/1 mm) < 80% and γ_(2%/2 mm) < 90%) was proposed and retrospectively validated in an independent cohort of 20 patients, comprising 618 treatment fractions.

RESULTS

Strong negative correlations were observed between ΔV and gamma indices, with γ_(1%/1 mm) showing the highest sensitivity (τ = -0.83), followed by γ_(2%/2 mm) (τ = -0.75). Volumetric variations exceeding 5% were associated with significant dosimetric degradation. The validation correctly identified all replanned patients and detected potential indicators of dosimetric divergence not recognized in routine clinical practice.

CONCLUSIONS

The thresholds γ_(1%/1 mm) < 80% and γ_(2%/2 mm) < 90%, combined with ΔV > 5%, proved effective for guiding replanning decisions. The proposed methodology demonstrated high sensitivity for detecting dosimetric deviations, with potential to enhance radiotherapy safety and effectiveness.

Для цитирования: Pereira da Silva Jayane Julia, Mendanha Ribeiro Caio Weber, Ludwig Jaime Luiz, Rocha Medrado Daniela, Melo da Silva Bruno Nunes, Alves Sarmento Ayane Cristine, de Moura Santos Edilmar. Dosimetric correlation between exit dose maps and anatomical changes in head and neck cancer treatment on Halcyon-E: A gamma analysis approach.. Journal of applied clinical medical physics. 2026. DOI: 10.1002/acm2.70710.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.