DOI 10.1016/j.clinbiomech.2017.01.005

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2017

Axial and torsional stiffness of pediatric prosthetic feet.

Taboga Paolo, Grabowski Alena M

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Prosthetic stiffness likely affects the walking biomechanics of toddlers and children with leg amputations, but the actual stiffness values for prostheses are not reported by manufacturers or in standardized testing procedures.

AIM: We measured axial (kA) and torsional (kT) stiffness from four brands of pediatric prosthetic feet (Trulife, Kingsley Mfg. Co., TRS Incorporated, and College Park Industries) over a range of foot sizes.

METHODS

We applied forces and torques onto prostheses with a materials testing machine that replicated those exhibited in vivo by using the kinetics measured from four non-amputee toddlers (2-3years) during walking.

FINDINGS: Across brands, kA averaged 35.2kN/m during heel loading, was more stiff during midfoot loading (121.8kN/m, P<0.001) and less stiff during forefoot loading (11.8kN/m, P=0.013). kA was similar across brands with no statistically significant effect of prosthetic foot size, with the exception of the TRS feet. Plantarflexion torsional stiffness (kT1), was not statistically different across brands. For every 1cm increase in foot size, kT1 increased 0.16kN·m/rad (P<0.001). College Park prostheses had 4.54kN·m/rad lower dorsiflexion torsional stiffness (kT2) (P<0.001) compared to other brands. For every 1cm increase in foot size, the kT2 applied on the foot increased 0.63kN·m/rad.

INTERPRETATION: The axial and torsional stiffness testing methods are reproducible and should be adopted by prosthetic foot manufacturers. Axial and torsional stiffness values of commercially available prosthetic feet should be publically reported to health practitioners to ensure evidence-based decisions and meet the specific needs of each patient with a leg amputation.

Для цитирования: Taboga Paolo, Grabowski Alena M. Axial and torsional stiffness of pediatric prosthetic feet.. Clinical biomechanics (Bristol, Avon). 2017. DOI: 10.1016/j.clinbiomech.2017.01.005.