MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/bpo.0000000000000194

Медицинская библиотека · Journal of pediatric orthopedics · 2014

Effect of asymmetrical configuration of pins in the TSF external fixator used for tibial lengthening in a pediatric population.

Zenios Michalis, Oyadiji S Olutunde

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The Taylor Spatial Frame (TSF) is a relatively new circular external fixator linked with a computer program used in limb reconstruction surgery. It is routinely used in our unit (Royal Manchester Children's Hospital) for pediatric tibial lengthening.

METHODS

The case notes and radiographs of 17 children who underwent 18 tibial lengthenings were reviewed. The TSF was attached to the tibia using a hybrid fixation method in the safe zone with two 1.8 mm tensioned smooth wires and 2 anteriorly placed hydroxyapatite-coated pins. Mechanical compression testing of our TSF construct was performed and compared with a double-ring Ilizarov frame construct fixed with only tensioned wires.

RESULTS

Out of the 18 tibiae lengthened 14 (77.8%) had a significant anterior angulation (>10 degrees) at the end of the distraction period that was corrected with a residual TSF program, with no known long-term consequences. The mechanical tests showed that the Ilizarov frame showed a more symmetrical distribution of loading in compression when compared with the TSF. With increasing load the TSF had the tendency to angulate anteriorly, that is, the apex of the angulation was pointing toward the 2 pins.

CONCLUSIONS

The problem of anterior angulation observed in our clinical and mechanical studies is due to instability of the fixation of the proximal part of the frame. Surgical techniques that can improve the stability of the proximal fixation and promote symmetrical loading are discussed.

Для цитирования: Zenios Michalis, Oyadiji S Olutunde. Effect of asymmetrical configuration of pins in the TSF external fixator used for tibial lengthening in a pediatric population.. Journal of pediatric orthopedics. 2014. DOI: 10.1097/bpo.0000000000000194.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.