Long echo time proton magnetic resonance spectroscopy for estimating relative measures of lipid unsaturation at 3 T.

Troitskaia Alice, Fallone B Gino, Yahya Atiyah

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To examine the behavior of lipid olefinic and diallylic resonances as a function of PRESS (point resolved spectroscopy) echo time (TE) to determine an optimal long TE value for their measurement at 3 T.

MATERIALS AND

METHODS

Experiments were conducted on nine oils (almond, canola, cod liver, corn, linseed, peanut, sesame, sunflower, and walnut oil) and on vertebral and tibial bone marrow in vivo at 3 T. The methylene (or methyl + methylene), diallylic, and olefinic resonances were measured with PRESS with multiple TEs.

RESULTS

J-coupling evolution effects on the olefinic and diallylic peaks appeared to be minimized when TE = 200 msec. The TE = 200 msec olefinic/methylene and diallylic/methylene peak area ratios calculated for each oil correlated well with ratios deduced from oil compositions in the literature (R(2) = 0.92 and 0.98 for the olefinic and diallylic protons, respectively). In addition, the relative amounts of bone marrow unsaturation of vertebral and tibial bone marrow inferred from the TE = 200 msec olefinic/(methyl + methylene) peak area ratio agreed with values estimated from the literature.

CONCLUSION

A PRESS sequence with a long TE value of 200 msec is suitable for determining relative amounts of lipid unsaturation at 3 T.

Для цитирования: Troitskaia Alice, Fallone B Gino, Yahya Atiyah. Long echo time proton magnetic resonance spectroscopy for estimating relative measures of lipid unsaturation at 3 T.. Journal of magnetic resonance imaging : JMRI. 2012. DOI: 10.1002/jmri.23868.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.