MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/s25030672

Медицинская библиотека · Sensors (Basel, Switzerland) · 2025

Improving Stroke Treatment Using Magnetic Nanoparticle Sensors to Monitor Brain Thrombus Extraction.

Jyoti Dhrubo, Reeves Daniel, Gordon-Wylie Scott, Eskey Clifford, Weaver John

СтатьяПеревод пока не подготовлен

Аннотация

(1)

Background

Mechanical thrombectomy (MT) successfully treats ischemic strokes by extracting the thrombus, or clot, using a stent retriever to pull it through the blood vessel. However, clot slippage and/or fragmentation can occur. Real-time feedback to a clinician about attachment between the stent and clot could enable more complete removal. We propose a system whereby antibody-targeted magnetic nanoparticles (NPs) are injected via a microcatheter to coat the clot, oscillating magnetic fields excite the particles, and a small coil attached to the catheter picks up a signal that determines the proximity of the clot to the stent. (2)

Methods

We used existing simulation code to model the signal from NPs distributed on a hemispherical clot with three orthogonally applied magnetic fields. An in vitro apparatus was built that applied fields and read out signals from a 1.5 mm pickup coil at a variable distance and orientation angle from a sample of 100 nm iron oxide core/shell NPs. (3)

Results

Our simulations suggest that the sum of the voltages induced in the pickup coil from three orthogonal applied fields could localize a clot to within 180 µm, regardless of the exact orientation of the pickup coil, with further precision added via rotation-correction formulae. Our experimental system validated simulations; we estimated an in vitro distance recovery precision of 41 µm with a pickup coil 1 mm from the clot. (4)

Conclusions

Magnetic NP sensing could be a safe and real-time method to estimate whether a clot is attached to the stent retriever during MT.

Для цитирования: Jyoti Dhrubo, Reeves Daniel, Gordon-Wylie Scott, Eskey Clifford, Weaver John. Improving Stroke Treatment Using Magnetic Nanoparticle Sensors to Monitor Brain Thrombus Extraction.. Sensors (Basel, Switzerland). 2025. DOI: 10.3390/s25030672.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.