Respiratory motion tracking of spine stereotactic radiotherapy in prone position.

Mizonobe Kazufusa, Akasaka Hiroaki, Uehara Kazuyuki, Oki Yuya, Nakayama Masao, Tamura Shuhei, Munetomo Yoshiki, Kubo Katsumaro, Kawaguchi Hiroki, Harada Aya, Mayahara Hiroshi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The CyberKnife system is a specialized device for non-coplanar irradiation; however, it possesses the geometric restriction that the beam cannot be irradiated from under the treatment couch. Prone positioning is expected to reduce the dose to normal lung tissue in spinal stereotactic body radiotherapy (SBRT) owing to the efficiency of beam arrangement; however, respiratory motion occurs. Therefore, the Xsight spine prone tracking (XSPT) system is used to reduce the effects of respiratory motion. The purpose of this study was to evaluate the motion-tracking error of the spine in the prone position.

MATERIALS AND

METHODS

Data from all 25 patients who underwent spinal SBRT at our institution between April 2020 and February 2022 using CyberKnife (VSI, version 11.1.0) with the XSPT tracking system were retrospectively analyzed using log files. The tumor motion, correlation, and prediction errors for each patient were examined. Furthermore, to assess the potential relationships between the parameters, the relationships between the tumor-motion amplitudes and correlation or prediction errors were investigated using linear regression.

RESULTS

The tumor-motion amplitudes in each direction were as follows: superior-inferior (SI), 0.51 ± 0.39 mm; left-right (LR), 0.37 ± 0.29 mm; and anterior-posterior (AP), 3.43 ± 1.63 mm. The overall mean correlation and prediction errors were 0.66 ± 0.48 mm and 0.06 ± 0.07 mm, respectively. The prediction errors were strongly correlated with the tumor-motion amplitudes, whereas the correlation errors were not.

CONCLUSIONS

This study demonstrated that the correlation error of spinal SBRT in the prone position is sufficiently small to be independent of the tumor-motion amplitude. Furthermore, the prediction error is small, contributing only slightly to the tracking error. These findings will improve the understanding of how to compensate for respiratory-motion uncertainty in the prone position.

Для цитирования: Mizonobe Kazufusa, Akasaka Hiroaki, Uehara Kazuyuki, Oki Yuya, Nakayama Masao, Tamura Shuhei, Munetomo Yoshiki, Kubo Katsumaro, Kawaguchi Hiroki, Harada Aya, Mayahara Hiroshi. Respiratory motion tracking of spine stereotactic radiotherapy in prone position.. Journal of applied clinical medical physics. 2023. DOI: 10.1002/acm2.13910.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.