Journal of applied clinical medical physics · 2023

Optimal threshold of a control parameter for tomotherapy respiratory tracking: A phantom study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sano Keisuke, Fujiwara Masayuki, Okada Wataru, Tanooka Masao, Takaki Haruyuki, Shibata Mayuri, Nakamura Kenji, Sakai Yusuke, Suzuki Hitomi, Takahashi Kanae, Tanaka Masahiro, Yamakado Koichiro

Аннотация

BACKGROUND: Radixact Synchrony® , a real-time motion tracking and compensating modality, is used for helical tomotherapy. Control parameters are used for the accurate application of irradiation. Radixact Synchrony® uses the potential difference, which is an index of the accuracy of the prediction model of target motion and is represented by a statistical prediction of the 3D distance error. Although there are several reports on Radixact Synchrony® , few have reported the appropriate settings of the potential difference threshold.

PURPOSE: This study aims to determine the optimal threshold of the potential difference of Radixact Synchrony® during respiratory tumor-motion-tracking irradiation.

METHODS: The relationship among the dosimetric accuracy, motion tracking accuracy, and control parameter was evaluated using a moving platform, a phantom with a basic respiratory model (the fourth power of a sinusoidal wave), and several irregular respiratory model waveforms. The dosimetric accuracy was evaluated by gamma analysis (3%, 1 mm, 10% dose threshold). The tracking accuracy was measured by the distance error of the difference between the tracked and driven positions of the phantom. The largest potential difference for 95% of treatment time was evaluated, and its correlation with the gamma-pass ratio and distance error was investigated. The optimal threshold of the potential difference was determined by receiver operating characteristic (ROC) analysis.

RESULTS: A linear correlation was identified between the potential difference and the gamma-pass ratio (R = -0.704). A linear correlation was also identified between the potential difference and distance error (R = 0.827). However, as the potential difference increased, it tended to underestimate the distance error. The ROC analysis revealed that the appropriate cutoff value of the potential difference was 3.05 mm.

CONCLUSION: The irradiation accuracy with motion tracking by Radixact Synchrony® could be predicted from the potential difference, and the threshold of the potential difference should be set to ∼3 mm.