Multi-wavelength pulse oximeter is not suitable for adjusting D(LCO) measurements.
Аннотация
BACKGROUND
Diffusing capacity of the lung for carbon monoxide (D(LCO)) can be affected by abnormal hemoglobin (Hb) or carboxyhemoglobin (COHb) levels. Predicted D(LCO) can be adjusted to reflect abnormal Hb or COHb levels. Until recently, blood sampling was required to determine Hb and COHb levels, but a new pulse oximeter, the Masimo RAD-57, can measure Hb and COHb noninvasively. We hypothesized that there would be no significant difference between the invasive and noninvasive Hb and COHb measurements for adjusting D(LCO).
METHODS
In patients referred to our university hospital for D(LCO) testing, we simultaneously took arterial blood gas samples and measured Hb and COHb with the RAD-57 (SpHb and SpCOHb, respectively). We analyzed the paired values and the Hb-adjusted and COHb-adjusted predicted D(LCO) values with t tests and Bland-Altman plots. We compared the differences in predicted D(LCO) to a clinical threshold of 3 mL/min/mm Hg.
RESULTS
SpHb differed from Hb measured via arterial blood analysis (12.1 ± 2.4 g/dL vs 13.3 ± 2.1 g/dL, P < .001). SpCOHb did not differ significantly from COHb (ie, measured via arterial blood analysis) (2.1 ± 4.0 vs 2.5 ± 2.3, P = .25), but there was wide variability. There were small but statistically significant differences in the adjusted predicted D(LCO), depending on whether blood or pulse oximetry values were used. Predicted D(LCO) adjusted for both Hb and COHb was 22.5 ± 4.8 mL/min/mm Hg measured with the RAD-57 and 23.5 ± 4.5 mL/min/mm Hg via arterial blood analysis (P < .001). The limits of agreement for pulse oximetry adjusted D(LCO) exceeded the clinical threshold of 3 mL/min/mm Hg for Hb adjustments and combined Hb + COHb. Predicted D(LCO) values differed by > 3 mL/min/mm Hg in 17% of patients.
CONCLUSIONS
Pulse oximetry may be of limited usefulness for adjusting either predicted or measured D(LCO) values, but might be useful to screen patients for invasive testing, particularly if the D(LCO) is close to the lower limit of normal.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.