MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejmp.2026.105879

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026

Evaluation of detector performance for bremsstrahlung measurement in total skin electron treatment.

Das Indra J, Dogan Serpil K

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Bremsstrahlung measurement in total skin electron treatment (TSET) is challenging due to low signal at extended distances (3-5 m). The suitability of various detectors for picocoulomb measurements to both electrons and bremsstrahlung are evaluated in this study.

METHODS

Multiple detectors, including a 0.6 cm3 Farmer chamber, Advanced Markus parallel-plate ionization chamber, electron diode, Edge detector, microSilicon, microDiamond, and EBT3 film were used for a 6 MeV high-dose-rate beam (2500 MU/min) TSET. Measurements were performed in solid water and a small water tank. Data were analyzed in terms of signal-to-noise ratio. For EBT3 film, appropriate scaling for exposure to electron and bremsstrahlung radiation, along with standard film-handling procedures, was applied.

RESULTS

Signal-to-noise performance with certain detector-electrometer combinations is critical for TSET measurement as shown for the Advanced Markus chamber and the microDiamond detector that exhibit signal saturation and not suitable in TSET. With higher signal fidelity detectors, we observed a more modest increase in bremsstrahlung dose with distance, typically ranging from 0.6 % to 2.0 %. An exponential relationship, Br = 0.67·e(0.0017·SSD), is provided to estimate the magnitude of bremsstrahlung within the limits of experimental accuracy.

CONCLUSIONS

The Farmer chamber, Edge detector, EFD diode, microSilicon detector, and EBT3 film are best suited for bremsstrahlung measurements. The increase in bremsstrahlung dose with distance is modest, and the proposed functional form provides a practical method for estimating its magnitude at any treatment distance.

Для цитирования: Das Indra J, Dogan Serpil K. Evaluation of detector performance for bremsstrahlung measurement in total skin electron treatment.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2026. DOI: 10.1016/j.ejmp.2026.105879.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.