m+Medelevator.RU
← Поиск публикаций
Medical physics · 2022

Radio-fluorogenic nanoclay gel dosimeters with reduced linear energy transfer dependence for carbon-ion beam radiotherapy.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Maeyama Takuya, Mochizuki Anri, Yoshida Kazuki, Fukunishi Nobuhisa, Ishikawa Kenichi L, Fukuda Shigekazu

Аннотация

PURPOSE: The precise assessment of the dose distribution of high linear energy transfer (LET) radiation remains a challenge, because the signal of most dosimeters will be saturated due to the high ionization density. Such measurements are particularly important for heavy-ion beam cancer therapy. On this basis, the present work examined the high LET effect associated with three-dimensional gel dosimetry based on radiation-induced chemical reactions. The purpose of this study was to create an ion beam radio-fluorogenic gel dosimeter with a reduced effect of LET.

METHODS: Nanoclay radio-fluorogenic gel (NC-RFG) dosimeters were prepared, typically containing 100 μM dihydrorhodamine 123 (DHR123) and 2.0 wt% nanoclay together with catalytic additives promoting Fenton or Fenton-like reactions. The radiological properties of NC-RFG dosimeters having different compositions in response to a carbon-ion beam were investigated using a fluorescence gel scanner.

RESULTS: An NC-RFG dosimeter capable of generating a fluorescence intensity distribution reflecting the carbon-ion beam dose profile was obtained. It was clarified that the reduction of the unfavorable LET dependence results from an acceleration of the reactions between DHR123 and H2 O2 , which is a molecular radiolysis product. The effects of varying the preparation conditions on the radiological properties of these gels were also examined. The optimum H2 O2 catalyst was determined to include 1 mM Fe3+ ions, and the addition of 100 mM pyridine was also found to increase the sensitivity.

CONCLUSIONS: This technique allows the first-ever evaluation of the depth-dose profile of a carbon-ion beam at typical therapeutic levels of several Gy without LET effect.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты