DOI 10.14744/anatoljcardiol.2019.01651
Determination of the effects of cryoablation for atrial fibrillation on esophageal functions.
Аннотация
OBJECTIVE
Periesophageal vagal plexus injury is a complication of cryoablation for atrial fibrillation (AF). The aim of this study is to investigate the effect of cryoablation on esophageal functions and to determine the relationship between esophageal temperature and esophageal motility.
METHODS
Twenty patients with symptomatic paroxysmal AF who underwent cryoablation were included in this study. The lowest cryoballoon temperature for each pulmonary vein (PV) was recorded. Esophageal temperature was measured using an esophageal probe during each cryoapplication. Esophageal manometry was performed before the procedure and one day after the procedure for each patient in order to assess the esophageal functions.
RESULTS
During the procedure, the highest esophageal temperature change was found in the left-side PVs in 13 patients (65%) and in the right-side PVs in seven patients (35%). No correlation was found between the lowest cryoballoon temperature and esophageal temperature change (r=0.22, p=0.05). It was detected that the lower esophageal sphincter pressure and esophageal contraction amplitude pressure decreased after the procedure (before: 19.7±9.3 mm Hg, after: 14.3±4.9 mm Hg, p=0.001; before: 84.5±28.3 mm Hg, after: 72.7±34.3 mm Hg, p=0.005, respectively). Five patients (25%) developed gastrointestinal symptoms after the procedure.
CONCLUSION
During cryoablation, esophageal temperature measurement can be performed to reduce the probability of esophageal injury. Cryoablation affects esophageal motility, and esophageal manometry can be performed to detect esophageal motility impairments in patients with gastrointestinal symptoms.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.