MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ultrasmedbio.2015.07.032

Медицинская библиотека · Ultrasound in medicine & biology · 2015

High-Accuracy Ultrasound Contrast Agent Detection Method for Diagnostic Ultrasound Imaging Systems.

Ito Koichi, Noro Kazumasa, Yanagisawa Yukari, Sakamoto Maya, Mori Shiro, Shiga Kiyoto, Kodama Tetsuya, Aoki Takafumi

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

An accurate method for detecting contrast agents using diagnostic ultrasound imaging systems is proposed. Contrast agents, such as microbubbles, passing through a blood vessel during ultrasound imaging are detected as blinking signals in the temporal axis, because their intensity value is constantly in motion. Ultrasound contrast agents are detected by evaluating the intensity variation of a pixel in the temporal axis. Conventional methods are based on simple subtraction of ultrasound images to detect ultrasound contrast agents. Even if the subject moves only slightly, a conventional detection method will introduce significant error. In contrast, the proposed technique employs spatiotemporal analysis of the pixel intensity variation over several frames. Experiments visualizing blood vessels in the mouse tail illustrated that the proposed method performs efficiently compared with conventional approaches. We also report that the new technique is useful for observing temporal changes in microvessel density in subiliac lymph nodes containing tumors. The results are compared with those of contrast-enhanced computed tomography.

Для цитирования: Ito Koichi, Noro Kazumasa, Yanagisawa Yukari, Sakamoto Maya, Mori Shiro, Shiga Kiyoto, Kodama Tetsuya, Aoki Takafumi. High-Accuracy Ultrasound Contrast Agent Detection Method for Diagnostic Ultrasound Imaging Systems.. Ultrasound in medicine & biology. 2015. DOI: 10.1016/j.ultrasmedbio.2015.07.032.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.