Real-world efficacy of MiniMed™780G recommended settings (glycemic target 100 mg/dL, active insulin time 2 hours) in youth and young adults with type 1 diabetes.

Bassi Marta, Spacco Giordano, Pezzotta Federico, Di Jorgi Margherita, Siri Giulia, Pintabona Andrea, Calevo Maria Grazia, Minuto Nicola, Maghnie Mohamad

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIMS: Despite growing evidence supporting the efficacy and safety of the MiniMed™ 780G recommended settings (Glucose Target 100 mg/dL and Active Insulin Time 2 hours), their adoption in routine practice remains limited, mainly due to concerns about hypoglycemia. This study aimed to evaluate the impact of switching to these settings in pediatric and young patients with type 1 diabetes (T1D).

METHODS

We conducted a retrospective longitudinal analysis in children and young adults using MiniMed™780G system at our center. Patients who switched from their initial settings to a glucose target of 100 mg/dL and an active insulin time of 2 hours for clinical indications were included. Data were retrospectively collected 3 months after switch. Glycemic metrics were compared over the 14 days before the switch (T0) and at 1 month (T1), and 3 months (T3).

RESULTS

Ninety-one patients with a mean age of 17.89y were included, 81.3% of whom already had a glucose target of 100 mg/dL at baseline. Therefore, in most cases the primary change was reducing AIT from 3 to 2 hours. After switching to the recommended settings, Time in Range (TIR) significantly increased (p<0.001) at T1 (71.9% vs 74.8%) and T3 (71.9% vs 75.0%). Time in target range (TITR) similarly improved from 47.2% at T0 to 51.4% at T1 and 50.9% at T3 (p<0.001) without any significant increase in time below range (TBR). The proportion of patients meeting all ADA-recommended glycemic targets rose from 29.5% at baseline to 40% at T3. Following the switch, the contribution of automatic correction boluses to the total insulin dose increased, while overall daily insulin requirements remained stable.

CONCLUSIONS

Switching to the recommended MiniMed™780G settings, driven primarily by AIT reduction in most patients, was safe and effective, improving glycemic control without increasing hypoglycemia. These findings support broader use of these settings in pediatric and young adult patient with type 1 diabetes.

Для цитирования: Bassi Marta, Spacco Giordano, Pezzotta Federico, Di Jorgi Margherita, Siri Giulia, Pintabona Andrea, Calevo Maria Grazia, Minuto Nicola, Maghnie Mohamad. Real-world efficacy of MiniMed™780G recommended settings (glycemic target 100 mg/dL, active insulin time 2 hours) in youth and young adults with type 1 diabetes.. Frontiers in endocrinology. 2025. DOI: 10.3389/fendo.2025.1670266.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.